Live Cell Imaging: Planning, Markers, and Applications

Live Cell Imaging: Planning, Markers, and Applications
paly

Live cell imaging provides direct spatial and temporal information for cell analysis. However, careful planning and marker selection are crucial for success. This article discusses the practical aspects of live cell imaging, including markers, fluorophores, and the microscope, as well as the benefits and applications of this technique.

  • Uploaded on | 0 Views
  • bronson bronson

About Live Cell Imaging: Planning, Markers, and Applications

PowerPoint presentation about 'Live Cell Imaging: Planning, Markers, and Applications'. This presentation describes the topic on Live cell imaging provides direct spatial and temporal information for cell analysis. However, careful planning and marker selection are crucial for success. This article discusses the practical aspects of live cell imaging, including markers, fluorophores, and the microscope, as well as the benefits and applications of this technique.. The key topics included in this slideshow are live cell imaging, cell analysis, markers, fluorophores, microscope,. Download this presentation absolutely free.

Presentation Transcript


1. Live cell imaging

2. Why live cell imaging? Live cell analysis provides direct spatial and temporal information Planning your experiment The markers/fluorophores The cells environment Practical aspects of the experiment: the microscope Photodamage Applications of live cell imaging

3. Select your markers carefully You only see a limited number of molecules/fluorophores 2 to3 channels in live cell imaging

4. Fluorophores Usually tag: GFP, mCherry, Venus, dTomato, etc Transient transfections Overexpression Inducible expression Endogenous levels of plasmid at endogenous promoter

5. What you need to do Keep the cells happy Optimize your experiment to get the most out of it Limit photodamage (cells will change their behavior)

6. Key components Preparation and holding of the cell specimen Temperature and CO 2 control Microscope Light: wavelength, intensity Image acquisition Type of live-cell imaging experiments

7. Unhappy cells

8. Contamination in cells will affect your experiment And Mycoplasma!

9. Media types in human cells Need FBS DMEM/RPMI: culture media, contains phenol red, which causes background fluorescence ! CO 2 -independent media for long experiments Leibowitz L15 media, no phenol red!

10. Holders Must have a #1.5 coverslip (0.17mm thick)

11. Maintaining live cells on the microscope Tight control of the environment is critical for successful live-cell imaging Heat within the specimen chamber or chamber holder Warm air stream over the stage Enclose the stage area/whole microscope Use CO 2 -independent media Use CO 2 source

12. Heated objectives Alternatively, need to heat the chamber and lense for 2-4hrs as lenses expand with heat Microscope also needs to be stable

13. Your microscope: temperature control Heat within the chamber holder Warm air stream over the stage Enclose the stage area Enclose the entire microscope

14. Your microsocope Active correction: Autofocus-Not ideal: extra light exposure and change plane in x, y, z Active Z position monitoring: Nikon and Zeiss Long term focus stability-important for time lapse work, not as important for short term observations with operator present

15. Perfect focus To overcome drift due to mechanical and thermal changes over time

16. Other features of microsocopes useful for cell imaging Keep the exposure constant Motorised stage to follow multiple cells (also need appropriate software) Shutter on illuminators so that the cells dont bleach

17. Photodamage Live cells poorly tolerate high exposure to light- true for transillumination and epifluorescence: cell death, compromised cell function and stress Targets: the cell, the medium, the fluorophore Generation of reactive oxygen species Blue light is very toxic to cells The longer the wavelength, the better You have to compromise!

18. Light flux at specimen Illumination system: 75W Xenon arc 490/10nm exciter filter (60%T) 505nm dichromatic mirror (85% reflectance) Flux at specimen: 380W/cm2 2500 times the flux of sunlight on the brightest day!

19. Minimize the exposure to the necessary for your experiment, not to make a pretty movie Kinetochore tracking in 3D 20 z-sections Every 7.5s seconds 5 minutes Thats a lot of exposure!

20. Minimum exposure to reduce photodamage Use a minimal exposure to maximize your data collection. Kinetochores are still there after 4min! Deconvolution (1cycle) can help restore your signal for presentation purposes.

21. Correcting for photobleaching

22. Type of live-cell imaging experiments one might do Time-lapse imaging (BF or TIRF) Photoactivated localized microscopy-PALM Fluorescence Recovery After Photobleaching-FRAP Fluorescence Correlation Spectroscopy-FCS Fluorescence Speckle Microscopy-FSM Fluorescence Resonance Energy Transfer-FRET

23. TIRF imaging of cells to image processes close to the membrane and focal adhesion

24. TIRF resolution in live-cell imaging 100-250nm in z-axis The evanescent field, resulting from total internal reflection of the beam excites fluorophores in a SMALL volume, close to the coverslip. Therefore sample photobleaching is very low

25. Fluorescence Recovery After Photobleaching- FRAP to look at 2D diffusion Very good for membrane dynamics

26. Photoactivation to determine movement of molecules and lifetime of subcellular structures

27. Fluorophore Photoconversion EosFP is a green fluorescent protein (emits at 516nm) from stony coral Near-UV radiation induces a conformational change in the protein Protein emission at 581nm Especially good for cell tracking in organisms

28. The birth of speckle microscopy

29. Fluorescence speckle microscopy to look at motion and turnover of macromoleulcar assemblies Courtesy of M. Mendosa/S. Besson

30. FSM gives information on flux and movement of actin during migration Courtesy of M. Mendosa/S. Besson

31. Quantitative analysis of FSM imaging gives information on actin movement during cell migration Courtesy of M. Mendosa/S. Besson

32. Fluorescence resonance energy transfer (FRET) FRET involves non-radiative energy transfer between donor and acceptor fluorophores Occurs over distances of 1-10 nm Emission and excitation spectrum must significantly overlap Can be used to measure close interaction between fluorophores and as a spectroscopic ruler to measure intermolecular distance Donor molecule Acceptor molecule Excitation Emission Excitation Emission FRET Intensity Wavelength

33. Example: the emission and absorption spectra of cyan fluorescent protein (CFP, the donor) and yellow fluorescent protein (YFP, the acceptor), respectively. CFP & YFP pair is currently the best for FP-based FRET. Fluorescence resonance energy transfer (FRET)

34. When to use FRET?

35. An Aurora B FRET probe as a tool to monitor differential phosphorylation FRET occurs when it is not phosphorylated Violin et al. 2003 Fuller et al. 2008 We;burn rt al, 2010 Donor Acceptor

36. Aurora B phosphorylation varies with substrate position Decreasing phosphorylation Michael Lampson, Dan Liu

37. Fluorescence resonance energy transfer (FRET) Intensity Wavelength Donor molecule Acceptor molecule Excitation Emission No FRET Intensity Wavelength Donor molecule Acceptor molecule Excitation Emission Emission FRET An important control in FRET studies is to photobleach the acceptor and demonstrate that donor emission does NOT decrease