PowerPoint presentation about 'Federal Aviation Administration Heat Flux Gauge Calibration Update Chapter HF Draft 2014 June Materials Meeting Switzerland Materials Working Group'. This presentation describes the topic on This document provides an agenda for the Federal Aviation Administration's Heat Flux Gauge Calibration meeting held in June 2014. The. The key topics included in this slideshow are . Download this presentation absolutely free.

Presentation Transcript


Slide1Federal AviationAdministration HEAT FLUX GAUGE CALIBRATION UPDATE (CHAPTER HF DRAFT) 2014 June Materials Meeting Switzerland Materials Working Group Michael Burns, FAA Tech Center June, 2014

Slide2Federal AviationAdministration Heat Flux Gauge Calibration June 2014 AGENDA • Calibration of Secondary Standard HFG • Coating • HFG Mounting • DAQ / Calibration Software • Calibration Procedure • Analysis • Reporting

Slide3Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Secondary Standard HFG • A minimum of 4 data points are required encompassing nominal heat flux values of 0.88, 1.76, 2.64 and 3.52 BTU/ft 2 *sec (1, 2, 3 and 4 W/cm 2 ). • The zero flux data point must be omitted. Zero offset data is not to be considered and data beyond these ranges must be disregarded.

Slide4Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Coating • The HFG must have a thin, full-faced, opaque coating of high quality, high-temperature, ultra-flat black paint having a diffuse surface finish. • The nominal dry film thickness of the coating on the sensor face must be 2 mils or less with the “orangey” copper face of the sensor not visible. Note: Mankiewicz Black Suede, Krylon Flat Black, Pyromark 1200 Flat Black paints or products of similar specification or function have been found suitable.

Slide5Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Data Acquisition System Specification • The calibration software must convert millivolt data of a Working HFG (gauge to be calibrated) to that of a Standardized gauge (having a known slope) in order to generate a calibration factor. • Calibration data between the range of 0.88 BTU/ft 2 *sec (1 W/cm 2 ) and 3.52 BTU/ft 2 *sec (4 W/cm 2 ) is used to linearize the slope. • Zero offset data is not to be considered and data beyond these two ranges must be omitted.

Slide6Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Mounting HFGs for Calibration • Each HFG must be mounted in a supporting device protruding through 0.5 inch (12.7 mm) thick rigid refractory board. • Each gauge has a 180º field of view of the radiant heat source therefore the surface must be mounted flush with the insulation board and not recessed. Note: Kaowool M-Board, Superwool 607 or product of similar specification or function has been found suitable.

Slide7Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Calibration Procedure STEP 2 Apply radiant heat to both sensors until the Standardized HFG reading reaches 4.4 BTU/ft 2 *sec (5 W/cm 2 ). A heat radiance increase from 0 to 4.4 BTU/ft 2 *sec (0 to 5 W/cm 2 ) must occur within 45 ± 5 seconds to minimize heating of surrounding materials and/or equipment. STEP 3 Immediately begin decreasing the radiant heat. A cool down rate from 4.4 BTU/ft 2 *sec (5 W/cm 2 ) to less than 0.88 BTU/ft 2 *sec (1 W/cm 2 ) must occur over a minimum time period of 90 seconds.

Slide8Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Calibration Procedure (Continued) STEP 4 At approximately 3.52 BTU/ft 2 *sec (4 W/cm 2 ) begin recording signal data continuously for both HFGs. STEP 5 At approximately 0.88 BTU/ft 2 *sec (1 W/cm 2 ) stop recording signal data.

Slide9Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Analysis • Multiply the Standardized HFG millivolt values by the known slope and calculate the heat flux level for each data point. A least squares fit procedure is performed using the calculated heat flux and the Working HFG millivolt values. The linearized slope is determined by: ∆Y (Units) / ∆X (mV) Where: Y = Standardized HFG (BTU’s/ft 2 *sec or W/cm 2 ) X = Working HFG (millivolts) Note: To convert BTU’s/ft 2 *sec to W/cm 2 : multiply slope by 1.135654 To convert W/cm 2  to BTU’s/ft 2 *sec: multiply slope by 0.88

Slide10Federal AviationAdministration Heat Flux Gauge Calibration June 2014 CHAPTER HF DRAFT – CALIBRATION Calibration Reporting Parameters

Slide11Federal AviationAdministration Heat Flux Gauge Calibration June 2014 TOPIC SUBTOPIC DETAILS Facility conducting calibration     Customer Lab performing future tests with the Working HFG   Calibration date Year / Month / Day   Calibration expiration date Year / Month / Day   Working HFG Information Manufacturer   Part number   Serial number   Gauge type Gardon Schmidt-Boelter Coating Paint Information Nominal Thickness (mils) Calibration Factor BTU/ft 2 *sec W/cm 2 Voltage / Resistance (Zero Flux) Pre Calibration Voltage Ohm Post Calibration Voltage Ohm Prior Calibration Data (If available) Date Calibration Factor BTU/ft 2 *sec W/cm 2 Facility where calibration was conducted Percent change in calibration factor Traceability records to a Standardized HFG

Slide12Federal AviationAdministration Heat Flux Gauge Calibration June 2014 TOPIC SUBTOPIC DETAILS Standardized HFG Information Manufacturer   Part number   Serial number   Gauge type Gardon Schmidt-Boelter Coating Paint information Nominal thickness (mils) Calibration Factor BTU/ft 2 *sec W/cm 2 Voltage / Resistance (Zero Flux) Pre Calibration Voltage Ohm Post Calibration Voltage Ohm Calibration Date of Standardized HFG Year / Month / Day Traceability Records to NIST Primary Standard HFG   HFG cooling water Temperature ̊ F ̊ C XY plot or raw data table     Linearity of HFG mV signals during calibration (Confidence level)     Calibration Traceability     Calibration Certificates of Devices (DAQ, N IST etc.  )  